User's Manual

Acclarix LX3 Series Diagnostic Ultrasound System User Manual Getting Started
- 29 -
5. To fix the battery, move the battery holder to the middle position (see the figure below).
6. Close the battery door and secure it.
WARNING
1. When the battery capacity is 20%, the battery status icon turns red.
2. When the battery capacity is 10%, the system displays a prompt “Low Battery. The system will
shutdown in 3 mins. Please plug in the power supply immediately."
To remove the battery:
1. Turn off and unplug the system.
2. Find the battery compartment, located at the lower right side of the system(see figure 3-1) .
3. Unscrew the two screws on the battery door, and open the battery compartment.
4. Move the battery holder (see the figure below) up or down, and remove two batteries respectively.