User's Manual
Acclarix LX3 Series Diagnostic Ultrasound System User Manual Specifications
- 163 -
A.7 Wi-Fi Specifications
Standard Conformance 802.11b, 802.11g ,802.11n
Frequency Band 2.412-2.484GHz(@2.4GHz band)
Modulation Technique
OFDM
DSSS
CCK
Typical Transmit
Power(±2dBm)
2.4G Transmit
1MbpsDSSS :17.3dBm
2MbpsDSSS :17.3dBm
5.5MbpsCCK :17.3dBm
11MbpsCCK :17.3dBm
6MbpsOFDM: 17.1dBm
9MbpsOFDM :17.1dBm
12MbpsOFDM: 17.1dBm
18MbpsOFDM :17.1dBm
24MbpsOFDM :16.2dBm
36MbpsOFDM :15.3dBm
48MbpsOFDM :14.6dBm
54MbpsOFDM: 13.8dBm
Wi-Fi Quality of Service
Data rate
802.11b: up to 11 Mbps @ 2.4 Ghz
802.11g: up to 54 Mbps @ 2.4 Ghz
802.11n: up to 65 Mbps @ 2.4 Ghz
Data security WPA2 encryption
Application-layer delay No requirement. It's not used in real time.
Application-layer
reliability
No requirement. Application failure will be notified to the user
immediately.
System capacity
No more than one device will be allowed to connect with the ultrasound
system.
System anti-interference Can be coexistent with other Wi-Fi devices.
Network interruption
alarm
Network interruption is notified by disconnection icon and failure in
transmission is notified in Transfer Status window.
EMC test process
Wi-Fi function is not affected when the system is imposed with radiation
interference complied with IEC60601-1-2:2014 standard.