SlideShow: Shows an imported collection of images slide by slide.
Export: Exports the imported images to USB storage device.
Delete: Deletes the imported images.
Figure 11-11 Demo Screen
11.3.5. Self-Check
The Self Check screen provides access to checking the fan status and to testing the LCD screen.
11.3.6. Tool
The Tool screen provides access to system maintenance which must be performed by authorized
service personnel. Please refer to the Service Manual for the details.
11.3.7. Transducer Element Check
Proper performance of the transducer is the prerequisites for acquiring images or signals that
provide the intended information for the users. To ensure the proper performance of a transducer, it
is suggested to implement the transducer element check each time a transducer is activated for use
or at regular transducer performance check.
To perform transducer element check:
1. Activate a transducer from the Transducer touch screen.
2. Open the Self-Check screen.
3. Press the Probe Check button, and the system will bring you a dialogue.
4. Ensure the residual coupling gel is removed from the transducer, and place the transducer on a
flat table without any object around the transducer. Then press OK on the dialogue to run the