Telemetry Transmitter User Manual Monitoring ECG
WARNING
When the PVCs Alarm is set to OFF, MFM-CMS won’t give an alarm prompt even if an
alarm occurs. In order to avoid endangering the patient’s life, the user should use this
function cautiously.
15.16.2.3 ARR Relearning
ARR relearning is operated on MFM-CMS.
Enter the Parameter Setup Window (way to entering refers to 8.2 Setting Parameters) > Click ECG ARR
on the left physiological parameter list > Click ECG Selflearn > Click Update Monitor to
confirm.
Pick this item to start a learning procedure, and ECG ARR LEARNING is displayed on the
screen. The ECG ARR LEARNING will start in the following status:
Connecting leads;
Starting ARR learning manually;
Switching calculation leads.
15.16.2.4 ARR Alarm Settings
ARR alarm settings are operated on MFM-CMS. Please refer to 8.2.1 Parameters Alarm Setting
for more alarm settings.
The users can switch on or off all arrhythmia alarms by ARR alarm settings. And some
arrhythmia alarms can be individually switched on or off. They are: R-ON-T, VT>2, COUPLET,
PVC, BIGEMINY, TRIGEMINY, TACHY, BRADY, MISSED BEATS, IRR, PNC, PNP,
VBRADY and VENT. Some arrhythmia alarms are preset to be on and cannot be turned off.
They are ASYSTOLE and VFIB/VTAC.