Telemetry Transmitter User Manual Monitoring ECG
ARR Types
Occurring Condition
PVC
Single PVC detected in normal heartbeats.
TACHY
Adult: RR interval for 5 consecutive QRS complex 0.5s.
Pediatric: RR interval for 5 consecutive QRS complex 0.375s.
BRADY
Adult: RR interval for 5 consecutive QRS complex ≥ 1.5s.
Pediatric: RR interval for 5 consecutive QRS complex ≥ 1s.
MISSED BEATS
If HR < 120 bpm, no beats are detected for
1.75 times average RR interval;
or if HR ≥ 120 bpm, no beats are detected for one second.
IRR
Consistently irregular heart rhythm
PNC
PACE NOT CAPTURE: no QRS complex detected in 300ms after a pace
pulse.
PNP
PACE
R NOT PACED: no pace pulse detected in 1.75 times RR interval
after a QRS complex.
VBRADY
VENTRICULAR BRADYCARDIA: Each RR interval for 5 consecutive
ventricular beats > 1000 ms.
VENT
VENTRICULAR RHYTHM: Each RR interval for 5
consecutive
ventricular beats ranges from 600 ms to 1000 ms.
15.16.2 ARR Analysis Menu
15.16.2.1 Switching ARR Analysis On and Off
Switching ARR on or off is operated on MFM-CMS.
Enter the Parameter Setup Window (way to entering refers to 8.2 Setting Parameters) > Click ECG on the
left physiological parameter list > Choose ARR Analysis, or click ECG ARR on the left
physiological parameter list to choose ARR Analysis > Click Update Monitor to confirm.
15.16.2.2 PVCs Alarm Settings
PVCs alarm settings are operated on MFM-CMS. Please refer to 8.2.1 Parameters Alarm Setting
for more alarm settings.
Select On in the menu to enable prompt message when an alarm occurs; select Off to disable the
alarm function, and there will be a symbol beside PVCs.
- 82 -