User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Monitoring ECG
ARR Types
Occurring Condition
PVC
Single PVC detected in normal heartbeats.
TACHY
Adult: RR interval for 5 consecutive QRS complex ≤ 0.5s.
Pediatric: RR interval for 5 consecutive QRS complex ≤ 0.375s.
BRADY
Adult: RR interval for 5 consecutive QRS complex ≥ 1.5s.
Pediatric: RR interval for 5 consecutive QRS complex ≥ 1s.
MISSED BEATS
If HR < 120 bpm, no beats are detected for
1.75 times average RR interval;
or if HR ≥ 120 bpm, no beats are detected for one second.
IRR
Consistently irregular heart rhythm
PNC
PACE NOT CAPTURE: no QRS complex detected in 300ms after a pace
pulse.
PNP
PACE
R NOT PACED: no pace pulse detected in 1.75 times RR interval
after a QRS complex.
VBRADY
VENTRICULAR BRADYCARDIA: Each RR interval for 5 consecutive
ventricular beats > 1000 ms.
VENT
VENTRICULAR RHYTHM: Each RR interval for 5
consecutive
ventricular beats ranges from 600 ms to 1000 ms.
15.16.2 ARR Analysis Menu
15.16.2.1 Switching ARR Analysis On and Off
Switching ARR on or off is operated on MFM-CMS.
Enter the Parameter Setup Window (way to entering refers to 8.2 Setting Parameters) > Click ECG on the
left physiological parameter list > Choose ARR Analysis, or click ECG ARR on the left
physiological parameter list to choose ARR Analysis > Click Update Monitor to confirm.
15.16.2.2 PVCs Alarm Settings
PVCs alarm settings are operated on MFM-CMS. Please refer to 8.2.1 Parameters Alarm Setting
for more alarm settings.
Select On in the menu to enable prompt message when an alarm occurs; select Off to disable the
alarm function, and there will be a symbol beside PVCs.
- 82 -