User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Monitoring ECG
To change the smart lead off setting, enter the Parameter Setup Window (way to entering refers to 8.2
Setting Parameters)
> Click ECG on the left physiological parameter list > Choose Smart
LeadOff > Click Update Monitor to confirm.
15.12 Setting Pace Status
Setting pace status is operated on MFM-CMS.
It is important to set the paced status correctly when you start monitoring ECG. To change the
paced status in the ECG Setup menu, enter the Parameter Setup Window (way to entering refers to 8.2
Setting Parameters)
> Click ECG on the left physiological parameter list > Choose Pace > Click
Update Monitor to confirm.
When Pace is set to On:
- Pace Pulse Rejection is switched on. This means that pacemaker pulses are not counted as
extra QRS complexes.
- Paced symbol is displayed as
|
on the main screen.
NOTE:
1 When monitoring a patient with a pacemaker, set Pace to On. If monitoring a patient
without a pacemaker, set Pace to Off.
2 If Pace is set to On, the system will not perform some types of ARR analysis.
WARNING
Some pace pulses can be difficult to reject. When this happens, the pulses are counted
as a QRS complex, and could result in an incorrect HR and failure to detect cardiac arrest
or some arrhythmias. Keep pacemaker patients under close observation.
15.13 ECG Calibration
ECG calibration is operated on MFM-CMS. This item is used to calibrate ECG waveform.
To calibrate ECG, enter the Parameter Setup Window (way to entering refers to 8.2 Setting Parameters) >
Click ECG on the left physiological parameter list > Click Calibration to calibrate ECG
waveform. Then click Stop to end calibration.
NOTE:
The device can’t be monitored during ECG calibration.
15.14 ECG Waveform Settings
ECG waveform setting is operated on MFM-CMS and is applicable to the wave on MFM-CMS.
User can select an appropriate setting. The bigger the value is, the wider the waveform is.
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s are optional. Please refer to 6.4.1 Setting Waveforms for
more information.