Telemetry Transmitter User Manual Monitoring ECG
AHA (American Standard) IEC (Europe Standard)
V1
Brown/ Red
C1
White/ Red
V2
Brown/ Yellow
C2
White/ Yellow
V3
Brown/ Green
C3
White/ Green
V4
Brown/Blue
C4
White/ Brown
V5
Brown/Orange
C5
White/ Black
V6
Brown/Purple
C6
White/ Purple
15.9.1 Electrode Placement for 3-lead
Take the American standard for example, see the following figure:
RA placement - directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
LL placement - on the left hypogastrium.
Electrode Placement for 3-lead
15.9.2 Electrode Placement for 5-lead
Take the American standard for example; see the following figure:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
RL placement: on the right hypogastrium.
LL placement: on the left hypogastrium.
V placement: on the chest, the position depends on your required lead selection.
- 76 -