Telemetry Transmitter User Manual Alarm Information
12.3 Prompts
Message Cause
ECG Arr Learning
The QRS template building required for
Arr. Analysis is in process.
SpO
2
Search Pulse
SpO
2
module is analyzing the patient
signal and searching for the pulse to
compute the saturation, when sensor is
connected with patient.
Weak Wi-Fi signal
The status that iT20 Wi-Fi signal is lower
than level 1, means Wi-Fi signal is weak.
12.4 Adjustable Range of Alarm Limits
ECG alarm limits are listed as follows: unit (bpm)
Patient Type
ALM HI
ALM LO
HR
ADU
300
15
PED
350
15
ST analysis alarm limits are listed as follows: unit (mV)
ALM HI
ALM LO
ST
2.0
-2.0
PVCs alarm upper limits are listed as follows:
ALM HI
ALM LO
PVCs
10
0
RESP alarm limits are listed as follows: unit (rpm)
Patient Type
ALM HI
ALM LO
RESP
ADU
120
6
PED
150
6