Telemetry Transmitter User Manual Alarm Information
Message Cause Alarm Level Action Taken
SpO
2
Low Perfusion
The pulse signal is too
weak or the perfusion of
the measurement site is too
low. The value displayed
may not be correct.
Low
Reconnect the SpO
2
sensor and change
the measurement
site. If problem
exists, please notify
biomedical engineer
or manufacturer’s
service staff.
SpO
2
Noisy Signal
There is interference with
SpO
2
measurement signals
and the waveform is
abnormal.
Low
Check the condition
of patient and avoid
patient movement;
make sure the cable
is well connected.
SpO
2
Light
Interference
Ambient light around the
sensor is too strong.
Low
Reduce interference
of the ambient light
and avoid sensor’s
exposure to strong
light.
Battery Low
Battery Low
High
Change the batteries
or
charge the
batteries.
No si
gnal for
telemetry transmitter
Telemetry transmitter didn’t
connect to MFM-CMS.
Medium
Check whether
network connection,
MFM-CMS and
wireless AP are work
normally, or contact
supplier.
Check the ECG
accessories
ECG cables have connected
to patient, and
the valid
authorized information for
ECG accessories
cannot be
detected.
Medium
Use the specified
ECG accessories.