User's Manual

Table Of Contents
Telemetry Transmitter User Manual Alarm Information
12.2 Technical Alarm Information
NOTE:
The ECG alarm information listed in the below table describes the lead names in
America. For the corresponding lead names in Europe, please refer to the section
Installing Electrodes.
Message Cause Alarm Level Action Taken
ECG Lead Off
1) The drive electrode or
more than one ECG limb
electrode falls off the skin;
2) ECG cables fall off the
telemetry transmitter.
Low
Make sure that all
electrodes, leads and
patient cables are
properly connected.
ECG LL Lead Off
ECG electrode LL falls off
the skin or the ECG cable
LL falls off the telemetry
transmitter.
Low
Make sure that all
electrodes, leads and
patient cables are
properly connected.
ECG LA Lead Off
ECG electrode LA falls off
the skin or the ECG cable
LA falls off the telemetry
transmitter.
Low
Make sure that all
electrodes, leads and
patient cables are
properly connected.
ECG RA Lead Off
ECG electrode RA falls off
the skin or the ECG cable
RA falls off the telemetry
transmitter.
Low
Make sure that all
electrodes, leads and
patient cables are
properly connected.
ECG V Lead Off
ECG electrode V falls off
the skin or the ECG cable V
falls off the telemetry
transmitter.
Low
Make sure that all
electrodes, leads and
patient cables are
properly connected.
ECG Signal Exceeded
ECG measuring signal is
beyond measuring range.
Low
Check lead
connection and
patient condition
ECG Noise
ECG measuring signal is
greatly interrupted.
Low
Check lead
connection and
patient condition
- 63 -