User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Alarm Information
Message Cause Alarm level
BRADY
Adult: RR interval for 5 consecutive QRS complex ≥
1.5s.
Pediatric: RR interval for 5 consecutive QRS complex
≥ 1s.
User-selectable
MISSED BEATS
If HR < 120 bpm, no beats are detected for 1.75 times
average RR interval; or if HR ≥ 120 bpm, no beats are
detected for one second.
User-selectable
IRR
Consistently irregular heart rhythm
User-selectable
PNC (with
pacemaker)
PACE NOT CAPTURE: no QRS complex detected in
300ms after a pace pulse.
User-selectable
PNP (with
pacemaker)
PACER NOT PACED: no pace pulse detected in 1.75
times RR interval after a QRS complex.
User-selectable
VBRADY
VENTRICULAR BRADYCARDIA: Each RR
interval for 5 consecutive ventricular beats > 1000 ms.
User-selectable
VENT
VENTRICULAR RHYTHM: Each RR interval for 5
consecutive ventricular beats ranges from 600 ms to
1000 ms.
User-selectable
RESP APNEA
RESP cannot be measured within the set apnea alarm
delay time.
High
(user-unselectable)
RR High RR measuring value is above upper alarm limit. User-selectable
RR Low RR measuring value is below lower alarm limit. User-selectable
SpO
2
High
SpO
2
measuring value is above upper alarm limit.
User-selectable
SpO
2
Low
SpO
2
measuring value is below lower alarm limit.
User-selectable
SpO
2
No Pulse
The signal of the measurement site is too weak, so the
telemetry transmitter can’t detect the pulse signal.
High
(user-unselectable)
PR High
PR measuring value is above upper alarm limit.
User-selectable
PR Low PR measuring value is below lower alarm limit. User-selectable
- 62 -