User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual System Setup
To clear the alarms latched, please choose Alarm Reset from the menu in the patient sector.
Refer to 6.3 Menu in the Patient Sector for details.
10.2.3 Date/Time Setup
To change date and time for MFM-CMS, please refer to 3.8 Setting Date and Time.
10.2.4 MFM-CMS System Alarm Setup
You can configure the alarm setting for the MFM-CMS.
You can set the duration for the audio pause to 1 minute, 2 minute or 3 minute. By click the
Audio Pause button on the main screen, you can activate or deactivate the audio pause function.
You can enable/disable alarm mute function by selecting/ deselecting the Alarm Mute check box.
When Alarm Mute is not selected, the alarm Mute is disabled. When Alarm Mute is selected,
the alarm mute function is enabled. The icon is displayed at the top area of the screen.
You can set alarm sound intervals by choosing the desired intervals from the drop-down list of
High Alarm Interval (s), Med Alarm Interval (s) and Low Alarm Interval(s).
NOTE:
Once a new alarm occurs, the system will neglect the existing settings of Alarm Mute and
generate a new alarm.
10.2.5 Changing Language
To change the display language, please select Main Screen > System Setup > User Maintain,
and input the correct password. Select Language Setup and select the desired language from the
drop-down list.
NOTE:
The user must restart MFM-CMS to make the change effective.
10.2.6 HL7
Users can set the interval for HL7 data to be sent and set the format of HL7 packing data sent by
MFM-CMS. The interval can be set to 1 to 120 minutes. HL7 data is sent in the format of HL7
Lower Level Protocol by default. If the item XML is selected, the data sent by MFM-CMS will
be packed in the XML format.
NOTE:
HL7 data is sent via the port 9100 by default.