User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual System Setup
for which you want to configure the module switch setting. On the right of the setup window,
configure the setting as required.
10.1.4 Help
Help information is available on this interface.
10.2 User Maintain
To access the settings interface of user maintain, you have to input a user password. The default
password is ABC.
10.2.1 Telemetry Transmitter Batch Settings
You can configure the alarm limit, alarm switch and alarm level for a group of telemetry
transmitters. You need to choose a patient type from Adult or Pediat before you configure the
alarm settings in Templet of Alarm Limit Adjusting Range. Choose the telemetry transmitters
needing to be configured in the right pane in which you may see a list of telemetry transmitters,
and click Config. The configuration in the left templet pane will be applied to the chosen
telemetry transmitters.
Besides, from the right pane, you can choose a telemetry transmitter whose alarm settings will
serve as the source of batch settings for other telemetry transmitters. Select one telemetry
transmitter in the right pane and click Obtain Monitor Configuration to obtain its parameter
alarm settings. The obtained configuration will be displayed in the left templet pane. Choose the
telemetry transmitters needing to be configured in the right pane and click Config to complete
batch settings.
NOTE:
1 If the patient type set in Templet of Alarm Limit Adjusting Range is different from
the one set on the telemetry transmitter to be configured, the system may fail to set
the configuration for telemetry transmitter.
2 The prompt message Success only indicates success in setting the configuration for
current activated parameters on the telemetry transmitter.
10.2.2 Telemetry Alarm Latch Setup
You can configure the alarm latching setting of the telemetry transmitter via the MFM-CMS.
Choose Main Screen > System Setup > User Maintain > Telemetry Alarm Latch Setup.
From the Telemetry Device List on the left of the setup window, choose the telemetry transmitter
for which you want to configure the latching setting. On the right of the setup window, configure
the setting as required.
The latching alarm setting of the telemetry transmitter is off by default. For telemetry transmitter
which has been offline and then online and in which the patient is admitted again, the latching
alarm configuration remains the same as the last configuration used by the telemetry transmitter.