User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual System Setup
Chapter 10 System Setup
The contents related to this chapter are all operated on MFM-CMS.
The System Setup function is used to modify the display information at the patient section
according to the real requirements. By using this function, you can observe the waveform,
parameter, and the parameter list as your desire. There are Common Setup, User Maintain, and
Factory Maintain to be set.
10.1 Common Setup
It is mainly used to make some conventional monitoring settings, such as Param Unit Setup,
Color Settings, Display Setup, Help and Telemetry Module Switch Setup.
10.1.1 Color Setup
The user can change the display color of all parameters and the other information of the
parameter is displayed as the same color. And the information includes waveform name, gain and
filter, real-time value (upper limit and lower limit), review waveform and so on. To change the
color of the parameter:
1. Please select Main Screen > System Setup > Common Setup > Color Setup. Click on the
Param Select to choose desired parameter, then choose desired color from the left color area
or input desired RGB values of red, green, blue directly.
2. After this, click on OK to confirm.
After setting, the color displayed on Color Setup colomn is the successfully chosen color. The
color on Initial Color colomn is set by default.
To get the default color, choose desired parameter from Param Select and click Default Settings,
and then click OK to confirm.
10.1.2 Display Setup
The user can set the bed numbers to be viewed on the screen. To change the display to be viewed,
please select Main Screen > System Setup > Common Setup > Display Setup and choose the
desired bed number 3, 4, 6, 8, 10, 12, 14, 16, 32, 64 from the drop-down list of display bed.
NOTE:
Two screens are needed if you want to simultaneously display the information of 64
telemetry transmitters.
10.1.3 Telemetry Module Switch Setup
You can configure the module switch setting of the telemetry transmitter via the MFM-CMS.
Choose Main Screen > System Setup > Common Setup > Telemetry Module Switch Setup.
From the Telemetry Device List on the left of the setup window, choose the telemetry transmitter
- 52 -