User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Setting Telemetry Transmitters via MFM-CMS
8.2.1 Parameters Alarm Setting
You can configure the alarm setting via the alarm display and configuration list on the parameter
setup window. You can configure the alarm switch, alarm level, alarm upper and lower limits.
The operation steps are shown as follows:
1. Choose a parameter from the physiological parameter list.
2. Configure the alarm settings in the alarm display and configuration list.
3. Click Update Monitor to bring the configuration into effect.
WARNING
1 Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
2 When the alarm is set to OFF, MFM-CMS won’t give an alarm even if an alarm
occurs. In order to avoid endangering the patient’s life, the user should use this
function cautiously.
3 Setting alarm limits to extreme values may cause the alarm system to become
ineffective. It is recommended to use the default settings.
8.2.2 Physiological Parameter Attribute and Configuration
For the telemetry transmitters, the MFM-CMS can display the parameter attributes of ECG, SpO
2
and RESP whose settings can also be configured.
You can configure the following items for the parameter ECG of the telemetry devices: lead type,
Calc. lead, pace, alarm source, filter, ECG Gain, ST Analysis, ARR Analysis, Smart LeadOff,
Hum Filter and electrode setup.
You can configure the following items for the parameter SpO
2
of the telemetry devices:
Sensitivity and Alarm Source (when choose PR).
You can configure the following items for the parameter RESP of the telemetry devices: RESP
Lead, Apnea Time and ECG Gain.
NOTE:
Telemetry transmitter reads leads style automatically. Electrode setup is only applicable
to ECG lead off alarms for telemetry transmitter.
- 46 -