User's Manual

Table Of Contents
Telemetry Transmitter User Manual Setting Telemetry Transmitters via MFM-CMS
8.2.1 Parameters Alarm Setting
You can configure the alarm setting via the alarm display and configuration list on the parameter
setup window. You can configure the alarm switch, alarm level, alarm upper and lower limits.
The operation steps are shown as follows:
1. Choose a parameter from the physiological parameter list.
2. Configure the alarm settings in the alarm display and configuration list.
3. Click Update Monitor to bring the configuration into effect.
WARNING
1 Prior to monitoring, make sure that the alarm limit settings are appropriate for your
patient.
2 When the alarm is set to OFF, MFM-CMS won’t give an alarm even if an alarm
occurs. In order to avoid endangering the patient’s life, the user should use this
function cautiously.
3 Setting alarm limits to extreme values may cause the alarm system to become
ineffective. It is recommended to use the default settings.
8.2.2 Physiological Parameter Attribute and Configuration
For the telemetry transmitters, the MFM-CMS can display the parameter attributes of ECG, SpO
2
and RESP whose settings can also be configured.
You can configure the following items for the parameter ECG of the telemetry devices: lead type,
Calc. lead, pace, alarm source, filter, ECG Gain, ST Analysis, ARR Analysis, Smart LeadOff,
Hum Filter and electrode setup.
You can configure the following items for the parameter SpO
2
of the telemetry devices:
Sensitivity and Alarm Source (when choose PR).
You can configure the following items for the parameter RESP of the telemetry devices: RESP
Lead, Apnea Time and ECG Gain.
NOTE:
Telemetry transmitter reads leads style automatically. Electrode setup is only applicable
to ECG lead off alarms for telemetry transmitter.
- 46 -