Chapter 4 Basic Operations
The general functions for telemetry transmitter are operated on MFM-CMS, which include
chapter 5 patient management, chapter 6 patient sector, chapter 7 viewing single bed, chapter 8
setting telemetry transmitter via MFM-CMS, chapter 9 review, chapter 10 system setup, chapter
11 alarm management, chapter 12 alarm information, chapter 13 printing, and chapter 14
database management.
The following introduction is about the basic operations for telemetry transmitter and
MFM-CMS.
4.1 Basic Operations for Telemetry Transmitter
4.1.1 Battery Installing and Replacing
Bottom View
Installing:
As pictured above, move the battery compartment latch right to open battery door. Install two AA
alkaline batteries following “+” or “-” indication, and then press the door to close correctly with
“ka-ka” sound.
Replacing:
Method is the same with installing.
Under the condition that telemetry transmitter continuously works at least for 15 minutes and
moreover it has a good Wi-Fi communication (typical network environment with no interference),
the telemetry transmitter can keep power on up to 20 seconds after taking batteries out.
Battery compartment latch
Green bar occurs under
locked status.
Red bar occurs under
unlocked status.