User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Default Settings
X stands for I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5 or V6.
ARR Analysis
ARR Analysis
On
PVCs Alarm Level
Medium
PVCs Alarm Switch
Off
ARR Alarm Settings
Alarm Switch
Alarm Level
Alarm Record
ASYSTOLE
On
High
Off
VFIB/VTAC
On
High
Off
R ON T
On
Medium
Off
VT > 2
On
Medium
Off
COUPLET
On
Medium
Off
PVC
On
Medium
Off
BIGEMINY
On
Medium
Off
TRIGEMINY
On
Medium
Off
TACHY
On
Medium
Off
BRADY
On
Medium
Off
MISSEDBEATS
On
Medium
Off
IRR
On
Medium
Off
PNC
On
Medium
Off
PNP
On
Medium
Off
VBRADY
On
Medium
Off
VENT
On
Medium
Off
C.4 RESP Default Settings
RESP Settings
ADU
PED
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit
30
30
Alarm Low Limit
8
8
Apnea Time
20s
Sweep
12.5mm/s
- 123 -