User's Manual

Table Of Contents
Telemetry Transmitter User Manual Product Specifications
Arrhythmia analyses
ASYSTOLE
VFIB/VTAC
COUPLET
VT>2
BIGEMINY
TRIGEMINY
VENT
R on T
PVC
TACHY
BRADY
MISSED BEATS
IRR
VBRADY
PNC
PNP
A.6 RESP
Method
Impedance RA-LL
Measurement lead
lead II
Calculation Type
Automatic
Respiration excitation waveform
Square Wave, 64 kHz(
±
10%), <500 μA
Measuring Sensitivity
Within baseline impedance range: 0.3 Ω
Waveform bandwidth
0.2 Hz to 2.5 Hz (-3 dB)
Baseline Impedance Range
200 Ω to 2500 Ω (leads cables 1 KΩ resistance)
RR Measuring Range
Adult
0 rpm to 120 rpm
Ped
0 rpm to 150 rpm
Resolution
1 rpm
Accuracy
Adult
6 to 120 rpm: ±2 rpm
0 to 5 rpm: not specified
Ped
6 to 150 rpm: ±2 rpm
0 to 5 rpm: not specified
Gain Selection (MFM-CMS) ×0.25, ×0.5, ×1, ×2, ×3, ×4, ×5
Sweep
(MFM-CMS) 6.25 mm/s, 12.5 mm/s, 25 mm/s
- 114 -