User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Product Specifications
Ventricular Rhythm
The interval of 5 consecutive ventricular beats ranges from 600
ms to 1000 ms
Ventricular Bradycardia The interval of 5 consecutive ventricular beats is more than 1000
ms
☆ Maximum start-up alarm time for Tachycardia
Ventricular Tachycardia
1 mV 206bpm
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
Ventricular Tachycardia
2 mV 195bpm
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
Response time of Heart
Rate Meter to Change in
HR
HR range: 80 bpm to 120 bpm
Within 11 seconds
HR range: 80 bpm ~ 40 bpm
Within 11 seconds
Tall T-wave Rejection Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17
minimum recommended 1.2 mV T-Wave amplitude
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
Complied with IEC 60601-2-27: 2011, Sect. 201.
7.9.2.9.101 b)
4), the HR value after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy: 80 bpm±1 bpm
Slow alternating ventricular bigeminy: 60 bpm±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm±1 bpm
Bidirectional systoles: 91 bpm±1 bpm
Time to Alarm for Heart
Rate alarm conditions
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
Complied with IEC 60601-2-27: 2011, Sect. 208.6.6.2.103.