Telemetry Transmitter User Manual Product Specifications
Ventricular Rhythm
The interval of 5 consecutive ventricular beats ranges from 600
ms to 1000 ms
Ventricular Bradycardia The interval of 5 consecutive ventricular beats is more than 1000
ms
Maximum start-up alarm time for Tachycardia
Ventricular Tachycardia
1 mV 206bpm
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
Ventricular Tachycardia
2 mV 195bpm
Gain 1.0: 10 s
Gain 0.5: 10 s
Gain 2.0: 10 s
Response time of Heart
Rate Meter to Change in
HR
HR range: 80 bpm to 120 bpm
Within 11 seconds
HR range: 80 bpm ~ 40 bpm
Within 11 seconds
Tall T-wave Rejection Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17
minimum recommended 1.2 mV T-Wave amplitude
Accuracy of Heart Rate
Meter and Response to
Irregular Rhythm
Complied with IEC 60601-2-27: 2011, Sect. 201.
7.9.2.9.101 b)
4), the HR value after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy: 80 bpm±1 bpm
Slow alternating ventricular bigeminy: 60 bpm±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm±1 bpm
Bidirectional systoles: 91 bpm±1 bpm
Time to Alarm for Heart
Rate alarm conditions
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
Complied with IEC 60601-2-27: 2011, Sect. 208.6.6.2.103.