Telemetry Transmitter User Manual Product Specifications
Pace Pulse
Pulse indicator Pulse is marked if the requirements of ANSI/AAMI
IEC 60601-2-27: 2011, Sect. 201.12.1.101.12 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.2 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
And
Amplitude: ±3 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
Pulse Rejection Pulse is rejected if the requirements of IEC 60601-2-
27: 2011,
Sect. 201.12.1.101.13 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to100 μs
Baseline Reset Time
< 3 s.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.11
Minimum input slew rate >2.5 V/S
Heart Rate
HR Calculation
Range
ADU: 15 bpm to 300 bpm
PED: 15 bpm to 350 bpm
Accuracy
±1% or ±1 bpm, whichever is greater
Resolution
1 bpm
Sensitivity
≥300 μV PP
QRS
QRS Detection Range
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.15.