User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Product Specifications
Pace Pulse
Pulse indicator Pulse is marked if the requirements of ANSI/AAMI
IEC 60601-2-27: 2011, Sect. 201.12.1.101.12 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.2 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
And
Amplitude: ±3 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to 100 μs
Pulse Rejection Pulse is rejected if the requirements of IEC 60601-2-
27: 2011,
Sect. 201.12.1.101.13 are met:
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 μs to100 μs
Baseline Reset Time
< 3 s.
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.11
Minimum input slew rate >2.5 V/S
Heart Rate
HR Calculation
☆ Range
ADU: 15 bpm to 300 bpm
PED: 15 bpm to 350 bpm
☆ Accuracy
±1% or ±1 bpm, whichever is greater
☆ Resolution
1 bpm
Sensitivity
≥300 μV PP
QRS
QRS Detection Range
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.15.
- 111 -