User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Product Specifications
☆ Sweep
MFM-CMS: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Telemetry Transmitter: 12.5 mm/s
☆ Bandwidth (-3 dB)
Diagnosis: 0.05 Hz to 150 Hz
Monitor: 0.5 Hz to 40 Hz
Surgery: 1 Hz to 20 Hz
☆
CMRR (Common
Mode Rejection Ratio)
Diagnosis: >95 dB
Monitor: >105 dB
Surgery: >105 dB
Notch In diagnosis, monitor and surgery modes: 50
Hz/60Hz (Notch
filter can be turned on or off manually)
☆
Differential Input
Impedance
>5 MΩ
☆ Input Signal Range
±10 mV PP
☆
Accuracy of Input
Signal Reproduction
The total error and frequency response comply with IEC
60601-2-27: 2011, Sect. 201.12.1.101.1.
☆
Electrode Offset
Potential Tolerance
±500 mV
Auxiliary Current (Leads
off detection)
Active electrode: <100 nA
Reference electrode: <900 nA
☆
Recovery time after
Defibrillation
<5 s
☆
Leakage current of
patient
<10 μA
☆ Scale signal
1 mVPP, accuracy is ±5%
☆ System noise
<30 μVPP
Multichannel Crosstalk
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.5.
Frequency and Impulse
Response
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.8.
- 110 -