User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Product Specifications
A.4.2 Software Performance
Trend
240-hour trend review for each telemetry transmitter
;
12-hour short trend dynamic display for each telemetry
transmitter;
Alarm events
720 pieces of parameter alarm events for each telemetry
transmitter
Alarm type
Physiological Alarm
Technical Alarm
Alarm mode
3 levels of audible and visual alarms
Nurse call records
(from t
he telemetry
transmitter
are stored by
MFM-CMS)
Store 100 groups
Patient call records
(from t
he telemetry
transmitter
are stored by
MFM-CMS)
Store 100 groups
Waveform storage and
review
96-hour waveform for each telemetry transmitter
A.5 ECG
Complies with IEC 60601-2-27: 2011
Lead Mode Automatic identification for leads
3-Lead: I, II, III
5-Lead: I, II, III, aVR, aVL, aVF, V
Support Smart Lead Off Check
(Telemetry transmitter) cannot choose calculation lead;
(MFM-CMS) supports choosing calculation lead.
Lead naming style AHA, IEC
☆ Display Sensitivity
MFM-CMS: 1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5
mm/mV (×0.5), 10 mm/mV (×1), 20 mm/mV (×2), 40 mm/mV
(×4), AUTO gain
Telemetry Transmitter: AUTO gain
- 109 -