User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Intended Use and Safety Guidance
Chapter 1 Intended Use and Safety Guidance
1.1 Intended Use/Indications for Use
Telemetry transmitter (hereinafter called iT20) must work with central monitoring system
(hereinafter called MFM-CMS) manufactured by EDAN.
Telemetry transmitter is intended to be used in clinical divisions of hospital environments,
including CCU and general wards (as Cardiology Dept., Respiratory Dept.). It is intended to be
used for adults and pediatrics. The monitored physiological parameters include: ECG, respiration
(RESP), oxygen saturation of arterial blood (SpO
2
) and pulse rate (PR).
1.2 Safety Guidance
Federal (U.S.) law restricts this device to sale by or on the order of a physician.
WARNING
1 Before using the device, the equipment, patient cable and electrodes etc. should be
checked. Replacement shall be taken if there is any evident defect or signs of aging
which may impair the safety or performance.
2 The electrodes expired are forbidden to be used.
3 Medical technical equipment such as telemetry monitoring system must only be used
by persons who have received adequate training in the use of such equipment and
who are capable of applying it properly. The user should have access to, and fully
read user manual (this book) before use. Harm to patient may occur if users’
operating is not in accordance with user manual.
4 It is prohibited that the operator touches battery and patient simultaneously.
5 Do not use the device with electrosurgical unit simultaneously.
6 EXPLOSION HAZARD-Do not use the device in a flammable atmosphere where
concentrations of flammable anesthetics or other materials may occur.
7 SHOCK HAZARD-To avoid the RISK of electric shock, MFM-CMS must only be
connected to a SUPPLY MAINS with protective earth. Never adapt the three-prong
plug from the MFM-CMS to fit a two-slot outlet.
8 Under simultaneous use of cardiac pacemaker and other patient-connected
equipment, the pacing impulse analysis function must be switched ON. Otherwise, the
pacing impulse may be counted as regular QRS complexes, which could prevent an
asystole event from being detected or could lead to false alarm of asystole.
9 Do not come into contact with the patient, table, or the telemetry transmitter during
defibrillation.
- 1 -