User's Manual
Table Of Contents
- Statement
- Responsibility of the Manufacturer
- Terms Used in this Manual
- Chapter 1 Intended Use and Safety Guidance
- Chapter 2 Overview
- Chapter 3 Installation of Telemetry Monitoring System
- Chapter 4 Basic Operations
- Chapter 5 Patient Management
- Chapter 6 Patient Sector
- Chapter 7 Viewing Single Bed
- Chapter 8 Setting Telemetry Transmitters via MFM-CMS
- Chapter 9 Review
- Chapter 10 System Setup
- Chapter 11 Alarm Management
- Chapter 12 Alarm Information
- Chapter 13 Printing
- Chapter 14 Database Management
- Chapter 15 Monitoring ECG
- 15.1 Overview
- 15.2 ECG Safety Information
- 15.3 ECG Display
- 15.4 Selecting Calculation Lead
- 15.5 Changing Size of ECG Waveform
- 15.6 Changing ECG Filter Settings
- 15.7 ECG Alarm Settings
- 15.8 Monitoring Procedure
- 15.9 Installing Electrodes
- 15.10 Setting Alarm Source
- 15.11 Smart Lead Off
- 15.12 Setting Pace Status
- 15.13 ECG Calibration
- 15.14 ECG Waveform Settings
- 15.15 ST Segment Monitoring
- 15.16 Arr. Monitoring
- Chapter 16 Monitoring RESP
- Chapter 17 Monitoring SpO2
- Chapter 18 Monitoring PR
- Chapter 19 Using Battery
- Chapter 20 Safety
- Chapter 21 Care and Cleaning
- Chapter 22 Maintenance
- Chapter 23 Warranty and Service
- Chapter 24 Accessories
- A Product Specifications
- B EMC Information
- C Default Settings
- D Abbreviation
Telemetry Transmitter User Manual Maintenance
Chapter 22 Maintenance
WARNING
1 Failure on the part of the responsible individual hospital or institution employing the
use of this equipment to implement a satisfactory maintenance schedule may cause
undue equipment failure and possible health hazards.
2 If you discover a problem with any of the equipment, contact your service personnel,
or your authorized supplier.
22.1 Inspecting
The overall check of the telemetry transmitter, including the safety check, should be performed
only by qualified personnel every 24 months, and each time after fix up.
The following items should be checked:
If the environment condition and power supply meet requirement.
If the power supply cord has damage and insulativity meets requirement.
If the device and accessories have damage.
Specified accessories.
If the alarm system can work properly.
If the printer can work properly and the paper meets the requirement.
Battery performance
If all monitoring functions are in good conditions.
If the grounding resistance and leakage current meet requirement.
If any damage or abnormality is found, please don’t use the telemetry transmitter and contact
local Customer Service Center.
22.2 Maintenance Task and Test Schedule
Maintenance shall be carried out at least once every two years, or as specified by local laws. The
following tasks are for EDAN-qualified service professionals only. Contact an EDAN-qualified
service provider if your device needs a safety or performance test. Clean and disinfect equipment
to decontaminate it before testing or maintaining it.
Maintenance and Test Schedule
Frequency
Safety checks. Selected tests on
the basis of IEC60601-1
At least once every two years, or as needed, after any
repairs where the power supply is removed or replaced,
or if the telemetry transmitter has been dropped.
Check all monitoring functions
and measuring functions
At least once every two years, or as needed.
- 100 -