iSE Series Electrocardiograph User Manual EMC Information
Electromagnetic Immunity
Guidance and manufacture’s declaration - electromagnetic immunity
The iSE is intended for use in the electromagnetic environment specified below. The
customer or the user of the iSE should assure that it is used in such an environment.
Immunity IEC/EN 60601 test Complianc Electromagnetic environment -
test level e level guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
iSE, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Conducted 3 V
rms
3V
rms
Recommended separation
RF distance
150 kHz to 80 MHz
150 kHz to
IEC/EN
61000-4-6
6Vrms
c)
in ISM
bands between
80 MHz
6Vrms
c)
in
d
1.2
P
150 kHz to 80
MHz
0.15 MHz and 80 ISM bands
MHz between
0.15 MHz
and 80
MHz
Radiated RF 3 V/m 3 V/m
d
1.2
P
80 MHz to 800
MHz IEC/EN 80 MHz to 2.7 GHz 80 MHz to
d
2.3
P
800 MHz to 2.7
GHz 61000-4-3 2.7 GHz
See Table 1. See Table
1.
d
6
P
/ E at RF
wireless
communications equipment bands
equipment (including peripherals
such as antenna cables and
external antennas) should be used
any part of the iSE, including cables
specified by the manufacturer).
(Portable RF communications