iSE Series Electrocardiograph User Manual EMC Information
The iSE is intended for use in the electromagnetic environment specified below. The
customer or the user of the iSE should assure that it is used in such an environment.
Immunity test
IEC/EN 60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
8 kV contact 8 kV contact Floors should be wood,
discharge (ESD) concrete or ceramic tile.
15 kV air 15 kV air
If floor are covered with
IEC/EN
synthetic material, the
61000-4-2
relative humidity should
be at least 30%.
Electrical fast
2 kV for power 2 kV for power
Mains power quality
transient/burst supply lines supply lines
should be that of a
typical commercial or
IEC/EN
61000-4-4
hospital environment.
Surge
1 kV line to line 1 kV line to line Mains power quality
should be that of a
IEC/EN
2 kV line to ground 2 kV line to
typical commercial or
61000-4-5 ground
hospital environment.
Power frequency 30 A/m 30 A/m Power frequency
magnetic fields should
(50 Hz / 60 Hz)
magnetic field
IEC/EN
61000-4-8
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.