Use Instructions
iSE Series Electrocardiograph User Manual EMC Information
Appendix 2 EMC Information
Electromagnetic Emissions
Guidance and manufacture’s declaration - electromagnetic emission
The iSE is intended for use in the electromagnetic environment specified below. The
customer or the user of the iSE should assure that it is used in such an environment.
Emission test Compliance
Electromagnetic environment -
guidance
The iSE uses RF energy only for its
RF emissions
CISPR 11
Group 1
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
Harmonic
emissions
IEC/EN 61000-3-2
Complies
The iSE is suitable for use in all
establishments, other than domestic and
those directly connected to the public
low-voltage power supply network that
Voltage supplies buildings used for domestic
fluctuations/ flicker
emissions
Complies
purposes.
IEC/EN 61000-3-3
NOTE:
The EMISSIONS characteristics of the system make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) the system might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Electromagnetic Immunity
Guidance and manufacture’s declaration - electromagnetic immunity
- 72 -