Use Instructions

iSE Series Electrocardiograph User Manual Cleaning, Care and Maintenance
2. Wipe the exterior of the device and accessories, including the main unit, DE18
sampling box, ECG cable, reusable electrodes and etc., using a soft cloth dampened
with diluted disinfectant.
3. Remove the remaining disinfectant if appropriate with a dry cloth.
4. Air dry the main unit/sampling box/ECG cable/reusable electrodes for at least 30
minutes in a ventilated and cool place.
The validated disinfectants for the main unit, sampling box and ECG cables are:
Ethanol (75%)
Isopropanol (70%)
The validated disinfectant for the reusable electrodes is:
Isopropanol (70%)
If Ethanol or Isopropanol is used for both cleaning and disinfecting, a new cloth is
required for the disinfection step.
CAUTION
1. Do not use high-temperature, high-pressure vapour or ionizing radiation as
disinfection methods.
2. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.
3. Clean and disinfect reusable electrodes after each use.
12.4 Care and Maintenance
EDAN will make available on request circuit diagrams, component part lists, descriptions,
or other information that will assist service personnel to repair those parts of the
equipment that are designated by EDAN as repairable by service personnel.
CAUTION
1. Besides the maintenance requirements recommended in this manual, comply with
local regulations on maintenance and measurement.
2. Do not open the casing. Once the device is opened, all warranties will lapse.
3. The maintenance operations like software upgrade of the device can only be
completed by EDAN-qualified service personnel.
4. Operate the cardiograph, charge or store the battery at a temperature of 40 (104ºF)
or lower. Exposure to higher temperature may reduce battery life, damage the battery,
and degrade cardiograph performance.
- 60 -