Use Instructions
iSE Series Electrocardiograph User Manual System Setting
Item Description
Axis calculation Choose from: Area Method or Amplitude Method
.
method
Tachycardia Enter a heart rate. The default value is 100 bpm.
(greater than)
Bradycardia (less Enter a heart rate. The default value is 60 bpm.
than)
P Wave Normal Choose from: 110ms or 120ms
.
Duration
When the P wave duration exceeds this setting, system will
prompt you P wave widens.
Algorithm Choose from: Normal or Weak.
Sensitivity
Serious Diseases Choose from: On, Off
.
Hint
When On, diagnosis with serious illnesses will be marked with
“►◄” in the ECG report.
RV5+SV1
If selected, this measuring parameter will be displayed in auto
measurements and ECG report. This is selected by default.
RV6/SV2
If selected, this measuring parameter will be displayed in auto
measurements and ECG report. This is not selected by default.
RR/PP
If selected, this measuring parameter will be displayed in auto
measurements and ECG report. This is not selected by default.
Maximum Signal
Choose from: ±5 mV
, ±10 mV, ±20 mV.
Range
Choose from: Bazzet
, Fridericia, Framingham, Hodges, and
QRS.
QTc Formulas
Bazzet, Fridericia, Framingham, Hodges, and QRS are displayed
in the ECG report as QTcBz, QTcFd, QTcFm, QTcHd, and
QTcQRS respectively.
Pacemaker
Choose from: Low, High
.
Detection Rate
10.7 User Setting
You have to enter password before configuring the user setting.