Use Instructions
iSE Series Electrocardiograph User Manual System Setting
the server.
Time Choose from: On
, Off.
Synchronization
When On, the system will send a synchronization request to the
server each time the electrocardiograph is powered on or retrieves
orders from server. And the time sent by the server is taken as the
current system time.
Encrypted Choose from: On
, Off.
Transmission
When On, the system will use encryption protocol during ECG
transmission. SFTP and SSL protocols will be employed for such
transmission. If you need to import another certificate for encrypted
transmission, please contact the manufacturer.
Receive push Choose from: On, Off
.
notifications
When On, the electrocardiograph can receive notifications of ECG
diagnosis from the server in real time.
When Off, the electrocardiograph will not receive such notifications.
MAC Address View the MAC address of the device.
Server Setting Order Server: set the address and port number of the server that
provides orders for the system. The port number allows a maximum
of five numbers for entry.
FTP or SFTP Server: set the address, port No., user name, and
password of the server that receives ECG files from the system.
The communication port allows a maximum of five numbers for
entry.
Connection test: checks if the server is connected.
NOTE:
For information on how to configure the network, consult your
network administrator.
Transmission Choose from: On
, Off.
Commitment
When On, the server will give a response whether it has received
ECG files or not to the system in each upload. This option is at the
bottom of the server setting mentioned above.
10.6 Parameter Setting