Use Instructions

iSE Series Electrocardiograph User Manual System Setting
Report addition to an automatic ECG printout.
VCG (visible when activated. Settings are displayed according to the functions
activated.)
Record Time Choose from: 10s, 1 min, 3 min, 5 min.
Report Type Choose from: VCG Report, TVCG Report, SAECG Report.
X/Y/Z Wave Choose from: Off, On.
If On, print X, Y, and Z waves on the VCG report.
Analysis Choose from: Off
, On.
If On, print the diagnosis on the VCG and SAECG report.
VCG Parameters Choose from: Off, On.
If On, print the detailed measurements on the VCG report.
QRS Gain Choose from: 10, 20
, 40, and 80 mm/mV.
Change the gain of QRS loop, the gain of P loop and T loop will be
synchronized.
SAECG Filter Choose from: 25-250 Hz, 40-250 Hz.
VCG Analysis Choose from: X, Y, Z.
Display the real-time heart of the lead you choose on the recording
screen.
HRV Report (visible when activated)
RR Waveform Choose from: Off, On
.
If On, print the compressed waveforms on the HRV report.
RR Interval List Choose from: Off
, On.
If On, print the RR interval list on the HRV report.
Pharma Study Record Info (visible when activated)
Auto Save Choose from: Off
, On.
If On, the system stores the ECGs recorded for the pharmaceutical
study.
Record Time Choose from: 0-1-3-5-10-15 (Minute)
, 0-1-3-5-7-9 (Minute),
0-2-4-6-8-10 (Minute), Customize.
This setting is related to the setting of report type below.
Report Type Choose from: All-Lead ECG Report, Single-Lead ECG Report
.
If All-Lead ECG Report is selected, the system will print out a 9-lead
or 12-lead ECG report at the set N minute.
If Single-Lead ECG Report is selected, the system will acquire 10s
single-lead waveforms at the
set N minute and at the end of
recording put all the acquired waveforms to a rhythm printout.
- 51 -