Use Instructions
iSE Series Electrocardiograph User Manual Preparing the Patient
Electrodes
IEC AHA
Locations
C7 V7
Left posterior axillary line on the same horizontal
level as C4 and C6
White/Orange Brown/Black
C8 V8
Left midscapular line on the same horizontal level
as C4 and C7
White/Blue Brown/Blue
C9
White/Yellow
V9
Brown/Yellow
Left paraspinal border on the same horizontal level
as C4 and C8
Attaching Electrodes
1. Ensure that the electrodes are clean.
2. For reusable electrodes:
Daub the electrode location with gel evenly. The electrode gel must cover an area the
size of the electrode but no larger, especially on the chest.
Apply electrodes to the prepared locations.
For disposable electrodes:
Apply electrodes to the prepared locations.
Clip or connect electrode adaptors to electrodes.
NOTE:
Long-time measurement with a strong negative pressure on the suction bulb may
cause reddening of the skin. When using the electrode on kids or patients with
delicate skin, squeeze the suction bulb lightly.
Proper skin preparation (abrasion if necessary), proper electrodes, and ECG cables
in fixed position are very important for a good signal quality.
CAUTION
The disposable electrodes can only be used for one time.