1.2 Safety Information
This section provides important safety information related to the use of iSE.
To use the system safely and effectively, firstly be familiar with the operation method of
Android and read the user manual in detail to be familiar with the proper operation so as
to avoid the possibility of system failure. The following warnings and cautions must be
paid more attention to during the operation of the system.
1.2.1 General Warnings
WARNING
1. The system is intended to be used by qualified physicians or personnel
professionally trained. They should be familiar with the contents of this user
manual before operation.
2. Only qualified service engineers can install this equipment, and only service
engineers authorized by the manufacturer can open the shell. Otherwise, safety
hazards may happen.
3. EXPLOSION HAZARD - Do not use the system in the presence of flammable
anesthetic mixtures with oxygen or other flammable agents.
4. Only the ECG cable and other accessories supplied by the manufacturer can be
used. Or else, the performance and electric shock protection cannot be
guaranteed. The system has been safety tested with the recommended
accessories, peripherals, and leads, and no hazard is found when the system is
operated with cardiac pacemakers or other stimulators.
5. Make sure that all electrodes are connected to the patient correctly before
operation.
6. Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrodes, do not come in contact with earth or any other
conducting objects.
7. Disposable electrodes must be used during defibrillation.
8. Electrodes of dissimilar metals should not be used; otherwise it may cause a
high polarization voltage.