12.9 NIBP Alarm
When NIBP Alarm is set to On, the physiology alarm occurs if any measurement value of
Systolic pressure, Mean pressure, Diastolic pressure exceeds alarm limit. The users can adjust the
alarm limit by accessing NIBP Setup > Alarm Setup > Sys Alarm/Map Alarm/Dia Alarm.
12.10 Resetting NIBP
When the pressure does not work properly and the system fails to give a message for the problem,
pick Reset in the User Maintain > NIBP Maintain menu to activate self-test procedure, and
thus restore the system from abnormal performance.
12.11 Calibrating NIBP
NIBP is not user-calibrated. Cuff-pressure transducers must be verified and calibrated on a yearly
interval by a qualified service professional. See the Service Manual for details.
12.12 Leak Test
This item is used for leak test. Turn the knob to pick the Leak Test item in the User Maintain >
NIBP Maintain menu to start the air leakage test. When the item is selected, it will change into
Stop. If this item is selected again, the system will stop air leakage test. And the item returns to
Leak Test.
WARNING
This pneumatic test other than being specified in the EN 1060-1 standard is to be used by
the user to simply determine whether there are air leaks in the NIBP airway. If at the end
of the test the system gives the prompt that the NIBP airway has air leaks, please contact
the manufacturer for repair.
12.12.1 Procedure of Leak Test
Connect the cuff securely with the socket for NIBP air hole.
Wrap the cuff around the cylinder of an appropriate size.
Access User Maintain > NIBP Maintain.
Turn the knob to the Leak Test item and press the item. Then the prompt of Leak
Testing will appear indicating that the system has started performing leak test.
For iM70/iM80 with the Omron module:
The system will automatically inflate the pneumatic system to about 180 mmHg
(285mmHg for Omron Module). After 20 seconds (4 minutes for Omron Module), the
system will automatically open the deflating valve, which marks the completion of a
pneumatic measurement.