Patient Monitor User Manual Alarm Information
- 47 -
Message Cause Alarm Level Action Taken
CO
2
Check Adapter
1) For the Respironics CO
2
module: The cannula is off
or disconnected.
2) For the EDAN CO
2
module: The water trap is
disconnected or not
properly connected.
Low
1) For the
Respironics CO
2
module: Check
whether the adapter
is properly connected
or replace the
adapter.
2) For the EDAN
CO
2
module:
Properly connect the
water trap.
AA Out Range
The AA concentration
exceeds the accuracy range
of AG module.
High
Reduce AA
concentration.
O
2
Sensor Error
The oxygen sensor of the
sidestream AG module has
a failure.
Medium
Stop measuring of
AG module, and
notify biomedical
engineer or
Manufacturer’s
service staff.
QuickTemp Comm
Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or manufacturer’s
service staff.
Temp exceed limit
The TEMP value is beyond
the range of +25°C ~
+45°C.
Med
Put the sensor into
the sensor bracket,
take it out and
measure again.
No Temp Sensor
TEMP sensor is not
connected to the TEMP
module.
Low
Connect the sensor
and the monitor well,
and measure again.
Ambient temp too
high
The Sensor temperature is
higher than +40°C.
Low
Put the sensor into
the sensor bracket,
measure again after
the ambient
temperature reaches
normal value.