User's Manual

Patient Monitor User Manual Alarm Information
- 45 -
Message Cause Alarm Level Action Taken
O
2
Out Of Range O
2
is out of range. High
Please make O
2
range resume
normal.
AG TEMP Out Of
Range
AG module temperature
out of range
High
Please make the
temperature resume
normal.
AG Baro Press Out
Of Range
AG module baro pressure
out of pressure
High
Please make the baro
pressure value
resume normal.
AG AA Id Unreliable
AG module can’t identify
the AG agent.
Medium
Reduce gas agent
type.
AG Span Calib In
Progress
AG module is calibrating. Low
Please wait
calibration finishing.
AG Calibration Fail
AG module calibration
failure
Medium
Please check whether
the module works
properly.
Unable To Calibrate
AG module can’t be
calibrated.
Medium
Please check whether
the module works
properly.
AG Zero In Progress AG module is zeroing. Low Please wait zeroing.
AG Occlusion
The sampling line is
clogged.
Medium
Replace the
sampling line.
AG Init Fail AG module has a failure. High
AG module works
improperly.
AG Data Limit Error AG module has a failure. High
AG module works
improperly.
AG Usa Error AG module has a failure. High
AG module works
improperly.
AG Cal Fail
AG module fails to
calibrate.
High
AG module works
improperly.
AG Zref Fail AG module fails to zero. High
AG module works
improperly.
AG Change Oxygen
Sensor
Replace oxygen sensor of
AG module.
Low
Please wait changing
finishing.
AG No Oxygen
Sensor
The oxygen sensor falls off
from the AG module.
High
Connect the sensor
again.