User's Manual

Patient Monitor User Manual Alarm Information
- 44 -
Message Cause Alarm Level Action Taken
C.O. Lack Param
C.O. measuring needs
parameters
High
Please input patient’s
height and weight.
AA Out Of Range
The AA concentration
exceeds the accuracy range
of AG module.
High
Reduce AA
concentration.
O
2
Sensor Error
The oxygen sensor of the
sidestream AG module has
a failure.
Medium
Stop measuring of
AG module, and
notify biomedical
engineer or
Manufacturer’s
service staff.
AG Comm Fail
AG module failure or
communication failure.
High
Stop measuring
function of AG
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
AG Zero Required AG module requires zero. Low
Please perform
zeroing.
AG Self-Testing AG module is self testing. Low
Please wait the self
testing finishing.
AG Replace O
2
Sensor
O
2
sensor needs to be
replaced.
High
Please replace the O
2
sensor.
AG Check Adapter AG module checks adapter. Low
Please wait check
finishing.
O
2
Cali Required O
2
needs to be calibrated. Low Please calibrate O
2
.
AG Software Error
AG module software
abnormal
High
Please replace
software revision.
AG Hardware Error
AG module has hardware
failure.
High
Please check whether
the hardware work
properly.
AG Motor Error
AG module motor
abnormal
High
Please check whether
the motor works
properly.
AG Uncalibrated AG module uncalibrated Low
Please calibrate the
AG module.
AG Replace Adapter
AG module needs to
change adapter.
High
Please replace the
adapter.