Patient Monitor User Manual Alarm Information
- 43 -
Message Cause Alarm Level Action Taken
TEMP T2 Sensor Off
Temperature cable of
TEMP channel 2 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected.
Excessive T1
TEMP1 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
Excessive T2
TEMP2 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
TEMP Comm Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
YY Sensor Off (YY
stands for the IBP
label name)
IBP sensor falls off. Low
Check the sensor
connection and
reconnect the sensor.
YY Comm Fail (YY
stands for the label
name)
IBP module failure or
communication failure
High
Stop measuring
function of IBP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
C.O. Comm Fail
C.O. module failure or
communication failure
High
Stop measuring of
C.O. module, or
notify biomedical
engineer or
Manufacturer’s
service staff.
C.O. TI No Sensor
C.O. TI sensor not
connected
Low
Insert injective
temperature sensor.
C.O. TB No Sensor
C.O. TB sensor not
connected
Low Insert TB sensor.
TEMP Out Of Range
TB measuring value is
above measuring range.
High
Please check TB
sensor.