User's Manual
Patient Monitor User Manual Default Settings
- 195 -
C.4 RESP
RESP Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 30 30 100
Alarm Low Limit 8 8 30
Apnea Time 20s
Calculation Type Auto
Resp Type II
Sweep 12.5mm/s
Amplitude 1
C.5 SpO
2
SpO
2
Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 100 100 95
Alarm Low Limit 90 90 88
Pitch Tone On
Sweep 12.5mm/s
C.6 PR
PR Settings ADU PED NEO
PR Source SpO
2
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 120 160 200
Alarm Low Limit 50 75 100