User's Manual
Patient Monitor User Manual Product Specification
- 186 -
N
2
O 0 to 100 vol% ±(2 vol% + 2% of reading)
HAL
ISO
ENF
0 to 8 vol%
8 to 25 vol%
±(0.15 vol% + 5% of reading)
Unspecified
SEV 0 to 10 vol%
10 to 25 vol%
±(0.15 vol% + 5% of reading)
Unspecified
DES 0 to 22 vol%
22 to 25 vol%
±(0.15 vol% + 5% of reading)
Unspecified
AwRR accuracy ±1rpm
Real-time gas
concentration
monitoring
Support
Zero Support
Work Mode Measurement
Apnea Alarm
Delay
10s, 15s, 20s, 25s, 30s, 35s, 40s; default value is 20s.
Interfering gas and vapour effects
CO
2
Gas or vapour Gas level
IRMA CO
2
IRMA AX+
Agents N
2
O
N
2
O
4)
60 vol%
_ 1&2) _ 1&2) _ 1) _ 1)
HAL
4)
4 vol%
_ 1) _ 1) _ 1) _ 1)
ENF, ISO, SEV
4)
5 vol% +8% of
reading
3)
_ 1) _ 1) _ 1)
DES
4)
15 vol% +12% of
reading
3)
_ 1) _ 1) _ 1)
Xe(Xenon)
4)
80 vol% -10% of
reading
3)
_ 1) _ 1)
He(Helium)
4)
50 vol% -6% of
reading
3)
_ 1) _ 1)
Metered dose
inhaler
propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH(Ethanol)
4)
0.3 vol%
_ 1) _ 1) _ 1) _ 1)
C
3
H
7
OH
(Isopropanol)
4)
0.5 vol%
_ 1) _ 1) _ 1) _ 1)