- 177 -
A.7 TEMP
Measurement method Thermal resistance
Channel 2
Sensor type YSI-10K and YSI-2.252K
Measuring Range
0
°C ~ 50 °C
Resolution
0.1
°C
Accuracy (Without sensor)
±0.1°C
Unit
℃, ℉
Refresh Time 1s ~ 2s
A.8 Quick TEMP
Measuring Range
25
°C ~ 45°C(monitoring mode)
35.5
℃~42℃(prediction mode)
Operating Temp
10
°C ~ 40°C
Sensor Type Oral/Axillary sensor, Rectal sensor
Resolution
0.1
°C
Accuracy(without sensor)
±0.1°C (25°C ~ 45°C) (monitoring mode)
Response time < 60s
Update time 1s ~ 2s
Warm-up time Less than 10 seconds
Prediction time Less than 30 seconds
A.9 IBP
Measurement method Direct invasive measurement
Channel iM80: 4 channels
iM50/iM60/iM70: 2 channels
Pressure sensor
Sensitivity 5 (μV/V/mmHg)
Impedance range 300 to 3000 Ω
Frequency response d.c. to 12.5 or 40 Hz
Zero
Range:
±200 mmHg