User's Manual

Patient Monitor User Manual Care and Cleaning
- 150 -
25.2.2 Cleaning the Accessories
25.2.2.1 Cleaning the ECG Cables and Lead Wires
NOTE:
1. Use only recommended cleaning substances and disinfectants listed in this document.
Others may cause damage (not covered by warranty), reduce product lifetime or
cause safety hazards.
2. Keep the cable and lead wires free of dust and dirt.
3. Never immerse or soak the ECG cable.
4. Inspect the cables after cleaning.
Recommended cleaning agents are:
Ethanol 75%
Isopropanol 70%
25.2.2.2 Cleaning the Blood Pressure Cuff
If the cover requires more rigorous cleaning, remove the air bladder first. Allow the cover to
thoroughly air dry before use.
Cuffs have been tested to withstand the following recommended cleaning agents: isopropyl 70%,
ethanol 75%.
25.2.2.3 Cleaning the SpO
2
Sensor
These reusable sensors should be cleaned and disinfected. The validated cleaning agents are listed
below:
Recommended cleaning agents are:
Ethanol 75%
Isopropanol 70%
25.2.2.4 Cleaning the IBP/C.O. Cables
NOTE:
1. Keep the cable free of dust and dirt.
2. Never immerse or soak the cable.
3. Inspect the cables after cleaning.
Recommended cleaning agents are:
Ethanol 75%
Isopropanol 70%