User's Manual

Patient Monitor User Manual Recording
- 142 -
4.
Pull about 2cm of the paper out, and close the recorder casing.
NOTE:
Be careful when inserting papers. Avoid damaging the thermo-sensitive print head.
Unless when inserting papers or shooting troubles, do not leave the recorder catch open.
22.5.5 Removing Paper Jam
When the recorder functions or sounds improperly, you should open the recorder casing to check
for a paper jam. Remove the paper jam in the following way:
Cut the record paper from the feeding edge.
Open the recorder casing.
Re-insert the paper.
NOTE:
If the monitor is not installed with a recorder, it will indicate RECORDER SETUP
NEEDED after pressing the Record button.