18.5 Setting Alarms
Here we take CO
2
alarm for example. This refers to CO
2
specific alarms. See the Alarms Chapter
for general alarm information. To change the alarm, please refer to the following steps:
1
Select the CO
2
Setup menu;
2
Select EtCO
2
Alarm High Limit or EtCO
2
Alarm Low Limit to adjust the alarm limit.
18.6 Setting Apnea Alarm Time
This determines the time limit after which the monitor gives an alarm if the patient stops
breathing.
1
Select the AG Setup menu to open it;
2
Select Apnea Alarm from the menu;
3
Choose the apnea alarm time from the pop-up list.
18.7 Working Status of ISA analyzer
Working status of the ISA analyzer can be indicated by the indicator. For the detailed information,
please refer to the following table.
Indication Status
Steady green light System OK
Blinking green light Zeroing in progress
Steady blue light Anesthetic agent present
Steady red light Sensor error
Blinking red light Check sampling line
18.8 Working Status of IRMA Module
The working status of the IRMA module can be transmitted by the IRMA probe. For the detailed
information, please refer to the following table.
Indication Status
Steady green light System OK
Blinking green light Zeroing in progress
Steady blue light Anesthetic agent present
Steady red light Sensor error
Blinking red light Check adapter