User's Manual

Patient Monitor User Manual Product Specification
- 183 -
Typical Rise Time CO
2
 200ms
HAL, ISO, ENF, SEV, DES  350ms
N
2
O  350ms
O
2
 450ms
Primary
Anaesthesia Agent
Threshold
 0.15 vol%
Second Anaesthesia
Agent Threshold
0.2 vol% + 10%
Agent
Identificaiton Time
< 20 seconds (typically < 10 seconds)
Total System
Response Time
< 3 seconds
Data Update Time 1 second
Accuracy(Standard Conditions)
GAS Measurement Range Accuracy
CO
2
0 to 15 vol%
15 to 25 vol%
±(0.2 vol% + 2% of reading)
Unspecified
N
2
O 0 to 100 vol% ±(2 vol% + 2% of reading)
HAL, ENF, ISO 0 to 8 vol %
8 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
SEV 0 to 10 vol %
10 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
DES 0 to 22 vol %
22 to 25 vol %
±(0.15 vol% + 5% of reading)
Unspecified
O
2
0 to 100 vol % ±(1 vol% + 2% of reading)
Apnea Alarm Delay 10s, 15s, 20s, 25s, 30s, 35s, 40s; default value is 20s.
Zero
Support
O
2
Compensation Support
N
2
O Compensation
Support
Interfering gas and vapor effects
CO
2
Gas or vapour Gas level
ISA CO
2
ISA AX+
Agents N
2
O