User's Manual

Patient Monitor User Manual Product Specification
- 180 -
FiCO
2
1mmHg
AwRR 1 rpm
r 2 mmHg, 0 to 40 mmHg
r 5 % of reading, 41 to 70 mmHg
r 8 % of reading, 71 to 100 mmHg
r 10 % of reading, 101 to 150 mmHg
EtCO
2
Accuracy
r 12 % of reading, RESP measurement value exceeds
80rpm (sidestream)
AwRR Accuracy
r 1 rpm
Sample Gas Flowrate
(sidestream)
50
fml/min
Stability
Short Term Drift Less than 0.8 mmHg over four hours
Long Term Drift Accuracy specification will be maintained over a 120 hour
period
O
2
Compensation
Range 0 ~ 100%
Resolution 1%
Default 16%
GAS Compensation
Range 0 ~ 20%
Resolution 0.1%
Default 0.0%
Zero Support
Work Mode Standby, Measurement
Barometric pressure
compensation
User setup
Balance gas compensation Including Helium, N
2
O and room air
Apnea Alarm Delay 10s, 15s, 20s, 25s, 30s, 35s, 40s; default value is 20s.
Interfering Gas Effect on EtCO
2
Measurement Values:
Gas or vapor Gas level (%) Quantitative effect/Comments
Nitrous oxide 60 Dry and Saturated Gas