User's Manual

Patient Monitor User Manual Product Specification
- 175 -
PR
Measurement range 40 bpm ~240bpm
Accuracy ±3bpm or 3.5%, whichever is greater
Omron Module
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring Interval in AUTO Mode 1/2/3/4/5/10/15/30/60/90 min, 2/4/8h
Continuous 5min, interval is 5s
Maximum measurement period Adult/ Pediatric: 160s
Neonatal: 80s
Adult/ Pediatric mode: 40bpm ~ 200bpm PR Measurement Range
Neonatal mode: 40 bpm ~ 240bpm
PR Accuracy
r 2 bpm or 2% of the readings
Measurement Range
Adult/ Pediatric Mode SYS: 60 mmHg ~ 250 mmHg
DIA: 40 mmHg ~ 200 mmHg
MAP: 45 mmHg ~ 235 mmHg
Neonatal Mode SYS: 40 mmHg ~ 120 mmHg
DIA: 20 mmHg ~ 90 mmHg
MAP: 30 mmHg ~ 100 mmHg
Cuff pressure measuring range 0 mmHg ~ 300 mmHg
Pressure Resolution 1mmHg
Accuracy
Maximum Mean Error ±5mmHg
Maximum Standard Deviation 8mmHg