User's Manual

Patient Monitor User Manual Product Specification
- 174 -
A.5 NIBP
EDAN Module
Measurement Method Oscillometric
Mode Manual, Auto, Continuous
Measuring interval in AUTO
Mode
1/2/3/4/5/10/15/30/60/90/120/240/480 min
Continuous 5min, interval is 5s
Measuring type SYS, DIA, MAP, PR
Measurement Range
Adult mode SYS: 40 mmHg ~ 270 mmHg
DIA: 10 mmHg ~ 215 mmHg
MAP: 20 mmHg ~ 235 mmHg
Pediatric mode SYS: 40 mmHg ~ 200 mmHg
DIA: 10 mmHg ~ 150 mmHg
MAP: 20 mmHg ~ 165 mmHg
Neonatal mode SYS: 40 mmHg ~ 135 mmHg
DIA: 10 mmHg ~ 100 mmHg
MAP: 20 mmHg ~ 110 mmHg
Cuff pressure measuring range 0 mmHg ~ 300 mmHg
Accuracy
Maximum mean error ±5mmHg
Maximum standard deviation 8mmHg
Pressure resolution 1mmHg
Maximum measuring period
Adult/Pediatric 120s
Neonate 90s
Typical measuring period 30s ~ 45s (depend on HR/motion disturbance)
Overpressure protection
Adult 297±3mmHg
Pediatric 240±3mmHg
Neonatal 147±3mmHg