User's Manual

Patient Monitor User Manual Recording
- 139 -
22.3 Recording Type
The monitor provides several types of stripe recording:
Continuous real-time recording
8 seconds real-time recording
Time recording
Alarm recording
Trend graph recording
Trend table recording
NIBP review recording
Arrhythmia review recording
Alarm review recording
Titration recording
Hemodynamic Calculation result recording
12-lead analysis recording
C.O. measurement recording
22.4 Starting and Stopping Recording
You can start and stop the recording in the following ways:
Continuous real-time recording Press the
Record button on the front panel to start the
recording, and repress it to stop the recording.
8 seconds real-time recording Record three waveforms selected in Recorder Setup menu
according to the setup time interval in
Recorder Setup
menu. It will automatically stop in 8 seconds.
Trend graph recording Enter the Menu > Review > Trend Graph menu, and press
the
Record button to start recording. Press the Record
button on the front panel to stop recording.
Trend table recording Enter the Menu > Review > Trend Table menu, and press
the
Record button to start recording. Press the Record
button on the front panel to stop recording.
NIBP review recording Enter the Menu > Review > NIBP Review menu, then
press the
Record button to start recording. Press the
Record button on the front panel to stop recording.
Arrhythmia review recording Enter the Menu > Review > ARR Review menu, and select
one arrhythmia alarm, then press the
Record button to start
recording. Press the
Record button on the front panel to
stop recording.