Use Instructions
Vital Signs Monitor User Manual Monitoring TEMP
- 69 -
Measure exposed skin:
Brush the hair and bangs aside if covering the area to be measured.
If an abnormally high or low reading is obtained:
Repeat the reading with the same Temp scanner; a correct reading will be reproducible.
Repeat the reading with another Temp scanner. Two Temp scanners with the same reading
will confirm the reading.
Sequential readings on the same patient in rapid succession will cool the skin; it is best to
wait about 30 seconds for the skin to recover from the cold probe.
Refer to the table below for Possible Causes of Abnormal TAT Temp Readings.
Type of Abnormal
Temperature
Possible Cause
Action Taken
Abnormally low
temperature
Dirty Lens.
Clean lens of scanner every two
weeks.
Releasing the button before
finished measuring.
Release the button after finished
measuring.
Measuring when an ice pack or
wet compress is on the
forehead.
Remove ice pack or wet compress,
wait 2 minutes, and re-take
temperature.